product
- Fluorescence Spectrometers
- Raman Microscopes
- Transient Absorption
- FTIR Spectrometers
- Lasers and LEDs
- UV-Vis Absorption
- Chromatography Data System
- Gas Chromatography
- Gas Chromatography-Mass Spectrometer (GC-MS)
- Liquid Chromatography
- Sampler and sample pretreatment system
- Column and Consumable
- Amino Acid Analyzer
- Isotope ratio mass spectrometry (IRMS)
- Ion-Molecule Reaction Mass Spectrometer(IMR-MS)
- Peripheral equipment and interface
- Ultra Centrifuge
- High Speed centrifuge
- Bench Top centrifuge
- DNA and RNA Purification
- Water Purification
- Pharmacy Refrigerator
- Ultra Low freezer
- Biomedical freezer
- Combined fridge-freezer
- Blood Bank Refrigerator
- Oven & Incubator
- Anaeraboic and Hyproxic Chambers
- Semi-Micro Balances
- Analytical Balances
- Precision Balances
- Industrial Scales
- Ash Analyzers
- Moisture Analyzers
- UV-Vis-NIR Microspectrophotometer
- Confocal Raman Micro spectrometer
- GeoImage Vitrinite Reflectance and Point Counting
- rIQ Glass Refractive Index Measuring System
- Standard and Software for Micro spectroscopy
- Bench Top Scanning electronic Microscope (SEM)
The DB30 UV-Vis Spectrophotometer is a powerful double beam instrument designed for precise and accurate measurement of absorption and transmission across various wavelengths, making it essential for analytical applications where precision is paramount.
Highlight:
◊ Variable bandpass options: 0.5, 1.0, 1.5, 2.0 or 4.0 nm over the range 190 nm – 1100 nm
◊ Fast scanning: up to 6,000 nm/min to aid sample analysis throughput
◊ Automatic accessory recognition: when connected to the DB30
◊ Optional Visacle® software: for easy PC control
◊ Touchscreen interface: intuitive menus and functions for ease of use for standalone control
◊ Multiple Data Storage Options: USB, SD card and internal data storage for convenient retrieval of methods and results
Engineered for versatility, this user-friendly spectrophotometer excels in a wide range of measurements, from standard absorbance in cuvettes to solid samples and flow cell experiments. Its high-throughput optical design delivers exceptional performance, ensuring superior stray light control, baseline stability, and unmatched wavelength and photometric accuracy.
The DB30 offers flexible operation, whether as a standalone device with its integrated touchscreen or through PC control using Visacle® software. Visacle provides a secure database for storing data and methods, and it can be upgraded to a CFR Part 11 compliant package for enhanced data protection and integrity.
Software:
As standard, the DB30 comes with a touchscreen for standalone operation controlling all modes of operation: photometry, time scan, wavelength scan, and data analysis, including calibration curves.
Additionally, Visacle® PC-controlled software enhances your DB30 experience by offering comprehensive control, validation, data analysis, and reporting—all in one streamlined package. With Visacle, your data is securely stored in an organised database, ready for analysis and easy export in .txt, .csv, or PDF formats.
For those needing extra security, Visacle is available as a CFR Part 11 compliant package. Key features include:
◊ Configurable user profiles with secure login and e-signatures
◊ Audit trail logging of all actions taken
◊ Permission system to restrict actions and data access to authorised users

Figure 1: Visacle Project Browser
Product Specifications:
| Parameter | Description |
|---|---|
| Optics | Czerny-Turner, Dual Beam Monochromator |
| Wavelength Range | 190 nm – 1100 nm |
| Spectral Bandwidth | 0.5 nm, 1 nm, 1.5 nm, 2 nm and 4 nm |
| Stray Light | ≤0.10% T (220 nm NaI, 340 nm NaNO2) |
| Wavelength Accuracy | ±0.1 nm D2 peak 656.1 nm |
| Wavelength Repeatability | ±0.1 nm |
| Photometric Range | Absorbance: -3.4 to +3.4, %T: 0 to 300, Concentration: 0,000 to 9,999 |
| Photometric Accuracy | ±0.002 Abs (0 – 0.5 Abs), ±0.004 Abs (0.5 – 1.0 Abs), ±0.008 Abs (1.0 – 2.0 Abs) ±0.3% T |
| Photometric Reproducibility | ±0.0003 Abs (0-0.5 Abs), ±0.0005 Abs (0.5-1 Abs), ±0.001 Abs (1.0-2 Abs) |
| Wavelength Scan Speed | 10, 100, 200, 400, 800, 1200, 2400, 3600, 6000 nm/min |
| RMS Noise | ≤0.00003 Abs (0 Abs, 500nm) |
| Baseline Stability | <0.0003 Abs/hr (500 nm, 2 hour lamp warm-up period) |
| Baseline Flatness | ±0.0009 Abs (200 nm – 950 nm) |
| Light Source | Tungsten-Halogen and Deuterium Lamps |
| Light Source Switching | Automatic switching selectable for 325 nm – 370 nm range |
| Detector | Silicon Photodiode |
| Display | 7” Touchscreen |
| Software | Touchscreen operation, PC-controlled software or CFR compliant PC-controlled software |
| Dimensions | 500 mm (W) × 475 mm (D) × 250 mm (H) |
| Net Weight | 20 kg (approx) |
| Power Supply | 100 – 240V, 50/60 Hz, 150VA |
| Ambient Temperature | 10°C – 35°C |
| Output Device | USB flash drive, SD card |
| Interface | USB PC interface |





73 Science Park Drive, #02-11 Cintech I, Singapore Science Park I, Singapore 118254
+65-6267 8921